Tilt-series preparation, alignment and tomogram reconstruction

Our initial steps for tilt-series motioncorrection, alignment and reconstruction are based on the tomotools package.

Motion Correction

  1. Generate a movies.star file that lists all frames:
     source params.sh
     cd $ROOT
     find frames/ > movies.star
    

    Open the movies.star file and paste the 3 first lines of the below cell on top, so that the final file looks like this:

     data_global 
     loop_ 
     _rlnMicrographMovieName 
     ./frames/Position_10_001_0.00_20241119_150207_EER.eer 
     ./frames/Position_10_002_3.00_20241119_150229_EER.eer 
     ./frames/Position_10_003_6.00_20241119_150252_EER.eer 
     ...
    
  2. Copy your gain reference file into $ROOT/frames/.

  3. Use the Relion command below to sum EER frames into fractions for the motion correction. Adapt the --eer_grouping parameter to specify the number of frames that will end up in one fraction. Ideally, you want a total of ~5 fractions with a total dose of ~150 e/A2.
     relion_convert_to_tiff --i ./movie.star --o frames_tiff/ --compression none --gain ./frames/*gain --eer_grouping 92  
    

    Move all files that ended up outside of the subfolder into ./frames_tiff.

  4. Modify the .mdoc file to update to .tif file extensions. You might first want to make a backup of the original mdoc:
     cp $TOMONAME.mrc.mdoc $TOMONAME.mrc.origin.mdoc
     sed -i 's/.eer/.tif/' $TOMONAME.mrc.mdoc
    
  5. Run the motion correction with tomotools

     conda activate tomotools
     # Set motioncor executable
     export MOTIONCOR_EXECUTABLE=$MOTIONCOR_EXECUTABLE
     # Motion / gain correction
     tomotools preprocess --mcbin 1 --frames ./frames_tiff/frames/ --gainref ./frames_tiff/gain-reference.mrc $TOMONAME.mrc ./ts-aligned/ 
    

Tilt-series alignment and initial tomogram reconstruction

  1. The tomotools reconstruct command will run a tilt-series alignment with AreTomo and a tomogram reconstruction in IMOD:

     tomotools reconstruct -d 3000 -b 4 --aretomo ./ts-aligned/$TOMONAME.mrc
    

    The resulting alignment and reconstruction can be used to check the quality of this tomogram.

WarpTools tomogram reconstruction

Tomograms that we selected for STA now need to be reconstructed a second time, using the previous AreTomo tilt-series alignment, and including a patch CTF estimation. For this, we use the WarpTools software.

  1. Export the tomotools alignment to warp
     cd $ROOT
     mkdir warp
     conda activate tomotools
    
     tomotools aretomo2warp --v2 --frames-dir ./frames_tiff ./ts-aligned ./
    
  2. With the shell script, reconstruct the tiltseries with warp
     cd $ROOT/warp
     $REPOSITORY/warp_reconstruction.sh $ANGPIX $FRAMEEXPOSURE $RECBINNING
    

Finally, the folder structure should look something like this:


Tomograms
└───Position_XX_X
    │   • Position_XX_X.mrc # Raw tilt series
    │   • Position_XX_X.mrc.mdoc # Mdoc file
    │
    └───frames # Subfolder for EER frames
    │       • Position_XX_X_001_-10.00_20251212_214747_EER.eer
    │       • Position_XX_X_002_-7.00_20251212_214827_EER.eer
    │       ...
    │       
    └───frames_tiff
    │   │   • gain-reference.mrc
    │   │
    │   └───frames # Subfolder for summed .tif frames
    │           • Position_XX_X_001_-10.00_20251212_214747_EER.tif
    │           • Position_XX_X_002_-7.00_20251212_214827_EER.tif
    │           ...
    │   
    └───ts-aligned # alignment and reconstruction by tomotools (AreTomo and IMOD)
    │   │   • Position_XX_X_ali_filtered_rec_bin_X.mrc # IMOD reconstruction 
    │   │   • Position_XX_X_ali.mrc # tilt series aligned with AreTomo
    │   │   ... # additional AreTomo alignment files, including even and odd tilt series
    │   │
    │   └───Position_XX_X_ali_Imod # Subfolder with Imod reconstruction files
    │         
    │
    └───warp # warp reconstruction
    │   │   • warp_frameseries.settings 
    │   │   • warp_tiltseries.settings    
    │   │
    │   └───frames  # raw frames
    │   └───imod    # imod alignment files 
    │   └───mdoc    
    │   └───processing   
    │       └───reconstruction
    │           │   • Position_XX_ali_X.XXApx.mrc # warp reconstruction (for picking)
    │           │   
    │           └───deconv
    │               │   • Position_XX_ali_X.XXApx.mrc # warp high-contrast reconstruction
    │

Improve tilt-series alignment with MissAlign

We have found the recently developed MissAlign software to drastically improve the quality of tilt-series alignments. Here, we show how to use MissAlign to improve the AreTomo alignment made with the previous steps. MissAlign is a neural network which can be trained on multiple tilt-series.

  1. Inside the Tomograms folder, generate a folder called miss-alignment and cd into it.

  2. Copy the update_warp_xml.py and config_template.yaml from MissAlign into this folder and update their content according to your specifications, as detailed in the MissAlign docs.

  3. For each tilt-series that you want to use for MissAlign training, copy its _ali.xml file (generated by warp) into the miss-alignment folder:
    cp ../Position_XX_X/warp/processing/Position_XX_X_ali.xml .
    
  4. Configure miss-alignment
    conda activate miss-alignment
    python update_warp_xml.py
    
  5. Run miss-alignment training
    miss-alignment train --config-file config_template.yaml --training-devices 0,1 --reconstruction-devices 2,2,2,2,2,3,3,3,3,3 --dataloaders-per-trainer 5 --start-at-iteration 0 --prepare-stacks $ANGPIX_REC
    

    This will take several hours, depending on your GPU configuration.

  6. Copy the _ali.xml files (which now contain the refined alignments) back into the warp processing folders:
    cp Position_XX_X_ali.xml ../Position_XX_X/warp/processing
    
  7. Reconstruct the tomogram again with WarpTools, now without first importing the alignments:
    cd $ROOT/warp
    WarpTools ts_defocus_hand --settings warp_tiltseries.settings --set_auto
    WarpTools ts_ctf --settings warp_tiltseries.settings --defocus_max 8 --range_high 4
    WarpTools ts_reconstruct --settings warp_tiltseries.settings --angpix $ANGPIX_REC --dont_invert --deconv
    